RCD test checker

Compare an Australian RCD or RCBO measured test row with the entered criterion and expected outcome.

  • Calculator
  • Protection
  • Australia
Choose a common device label, or select Custom for a project-specific test-record label.
mA
Enter the device residual current rating used for the test row.
Choose the measured row being compared with the entered criterion.
Choose the comparison basis for the rated-current trip-time row.
Choose the recorded result for the selected row.
ms
Enter the recorded trip time when the row tripped.
Itest = I_dn x m; margin_ms = tmax_ms - tmeas_ms; utilisation_% = tmeas_ms / tmax_ms x 100
  • Non-operating rows use the entered no-trip duration instead of a trip-time comparison.
  • Push-button rows use the recorded function outcome rather than a timed current-injection comparison.
  • The criterion is entered or selected by the user and must be checked against the project basis, current requirements and manufacturer data.
Formula variables
VariableMeaningUnitUse
I_dnRated residual currentmADevice rating entered from the RCD, RCBO or test record.
mTest multiplierxSelected current-injection row multiplier.
ItestTest currentmARated residual current multiplied by the selected row multiplier.
tmax_msMaximum trip timemsSelected or manually entered maximum time for a timed row.
tmeas_msMeasured trip timemsTrip time recorded by the test instrument for a timed row.
tnon_sNo-trip durationsEntered duration for a non-operating-current row.
margin_msTime marginmsDifference between the entered maximum time and measured trip time.
utilisation_%Time utilisation%Measured trip time as a share of the entered maximum time.
outcomeObserved outcome-Recorded trip, no-trip, operated or did-not-operate outcome.
More

RCD test checker technical guide

Compare a measured RCD or RCBO test row with the entered criterion and expected outcome.

Use this page after an RCD, RCBO or safety switch test has already been taken and one line of the record needs a disciplined check. The practical problem is not usually the multiplication. The problem is whether the row says enough for another competent person to see what was tested, which criterion was used and what still needs follow-up.

The calculator compares one row at a time. That is deliberate. RCD records become weak when a half-current no-trip check, a timed current-injection check and a push-button function check are collapsed into a single note such as "RCD OK". A useful row identifies the device, the test row, the selected or manually entered criterion, the observed outcome and the measured time where a timed row applies. This page turns those values into a consistent worksheet result without claiming that the device, circuit or installation has been fully verified.

The safest way to use the page is to treat the output as a record-checking aid beside the original instrument result, switchboard schedule, asset register or commissioning sheet. It helps expose missing values and marginal timing, but it does not decide the legal test interval, installation compliance, product suitability or final defect action.

Field use cases

Use the checker when the result already exists and the next job is to make the record usable. It suits electricians checking a test sheet before it leaves site, contractors preparing a handover set, estimators pricing a maintenance allowance, supervisors checking apprentice records and facility teams comparing current and previous register rows.

Field use cases
Work settingPractical questionUseful output
Switchboard handoverCan this timed row be read consistently by the next electrician or reviewer?Device reference, row, criterion, measured time, margin and status.
Temporary board registerDid the non-operating-current row stay closed for the recorded duration?Test current, no-trip duration and unexpected-trip warning where relevant.
Facilities maintenanceIs this entry comparable with the previous register entry for the same device?Same device label, same row wording and a clear entered basis.
Contractor supervisionHas a push-button function check been mixed with a timed current-injection row?Separate function-check status from timed trip-time rows.
Defect follow-upIs the concern late operation, no operation, or operation when it should not occur?A status that points to retest, investigation or escalation.

The calculator is not a schedule builder. It does not decide how often a workplace item must be tested, which devices must be installed, whether an installation is compliant, or whether a device should be replaced. It gives a clean reading of one row so the record can be discussed without ambiguity.

Recording one row

Start from the record in front of you, not from the result you hope to see. For a current-injection row, record the device label, rated residual current, selected row, observed trip or no-trip outcome, measured trip time where a trip is recorded, and the source of the comparison basis. For a push-button row, keep it as a function check. Do not convert it into a timed trip record.

Row record checklist
Record fieldStrong recordWeak record
Device referenceMatches the board schedule, circuit list, safety switch label, asset register or test sheet.Generic `RCD` label with no board, circuit or asset context.
Rated residual currentStates the mA rating used for a current-injection row.Missing rating on a 0.5x, 1x or 5x row.
Test rowSeparates non-operating current, rated-current trip time, five-times trip time and push-button function.Uses `tested` or `OK` without saying which row was performed.
Criterion basisStates Type 1, Type 2 or manual maximum where a timed row is compared.Uses a time value without saying where the basis came from.
Observed outcomeRecords trip, no trip, operated or did not operate.Leaves the reader to infer the outcome from a tick or short note.
Measured timeIncludes trip time where a timed row recorded a trip.Says `passed` or `failed` without the measured value.
Source recordKeeps the instrument record, site procedure, manufacturer basis or competent-person instruction with the job pack.Leaves the next reviewer to infer the method or threshold.

Test row interpretation matrix

The selected row changes what a good record needs to contain. A non-operating-current row is a no-trip check for the entered duration. A rated-current row needs a time basis. A five-times row should state the entered maximum used for that row. A push-button row records the inbuilt test facility only and should not be treated as a timed current-injection result.

Test row interpretation matrix
Selected rowExpected recordCalculator comparisonMain follow-up if flagged
Non-operating currentDevice should not trip for the entered duration.Checks the 0.5x test current and observed no-trip outcome.Confirm test point, loads, method and device condition if a trip is recorded.
Rated-current trip timeDevice should trip within the entered Type 1, Type 2 or manual basis.Compares measured time with selected or entered maximum time.Review criterion basis, instrument record and device data if late or no trip.
Five-times trip timeDevice should be compared with the entered maximum for the five-times row.Uses the manual maximum time entered for the row.Confirm the row basis before treating a fast-row result as acceptable.
Push-button functionInbuilt test facility should operate.Checks function outcome only.Treat a failure as a device or procedure review item, not a timed-row result.

This separation matters in real records. If a push-button result is filed as though it were a timed current-injection test, a future reviewer may assume a measurement exists when it does not. If a non-operating row is recorded as simply "no trip" without the duration and device label, the entry is hard to compare at the next inspection.

Choosing the entered criterion

The calculator only compares against the criterion entered or selected by the user. That is intentional. It avoids hiding a legal or product requirement inside the tool. Before using a preset or manual value, identify why that value belongs to the row: current standard, workplace procedure, project specification, equipment manual, regulator guidance or a competent-person instruction.

Use a preset only when it matches the row being checked. Use the manual maximum where the record, procedure or manufacturer data calls for a different timed-row basis. For a five-times row, the calculator expects a manual maximum because this page does not supply a universal hidden threshold. For a non-operating-current row, the important entered value is the duration of the no-trip check, not a trip-time maximum.

Criterion selection guide
SituationBetter calculator settingCheck before relying on it
Test sheet already states a Type 1 or Type 2 timed-row basisSelect the matching preset for the rated-current trip-time row.Confirm the device and row really match that basis.
Procedure gives a site-specific timed maximumUse manual maximum.Keep the procedure, manufacturer note or instruction with the record.
Five-times current-injection rowUse five-times row and manual maximum.Confirm the entered maximum is the one intended for that row.
Half-current no-trip rowUse non-operating current and enter the no-trip duration.Do not add a trip-time criterion unless the device trips unexpectedly.
Push-button checkUse push-button function.Keep it separate from timed current-injection evidence.

Record wording matrix

A row can look acceptable arithmetically and still be weak as a record. The useful output is a line another competent person can trace back to the same device, row and basis.

Bad row to usable row
Weak entryBetter worksheet wordingWhy it matters
`RCD OK``DB1 RCD-1, 30 mA, rated-current row, Type 2 300 ms basis, tripped 28 ms.`Identifies the device, row, basis and measured value.
`No trip``DB-TEMP RCD-2, 30 mA, non-operating row, 15 mA for 5 s, no trip.`Shows the half-current row and duration instead of a bare outcome.
`Push button OK``RCD-5 push-button function operated; no timed current-injection value recorded on this row.`Keeps function evidence separate from timed testing evidence.
`Failed``RCD-3 five-times row, manual 40 ms basis, no trip recorded; retest and device review remain open under the site procedure.`Names the row and the open action instead of hiding the failure mode.

Interpreting margin and utilisation

Margin and utilisation are record-review values, not a final decision about the device or installation. A positive margin says the timed value is below the entered maximum for that row. A negative margin says the measured time is above the entered maximum. A small margin still deserves attention when the basis, test point or device condition is uncertain.

Do not soften a late row because the device "still tripped". Marginal or late timing can point to a wrong criterion, transcription error, test point issue, device condition, connected load effect or method problem that needs follow-up.

Margin use guide
Calculator outputPractical readingSensible record action
Positive margin with clear device and criterionThe timed value is within the entered basis.Keep the row with the original instrument record and project basis.
Very small positive marginThe row is close to the entered maximum.Check data entry, instrument record, device type and whether the criterion is the intended one.
Negative marginMeasured time is above the entered maximum.Do not soften the result by changing the basis. Retest or investigate under the project procedure.
No margin shownThe selected row is non-operating, push-button or a no-trip timed failure.Use the status and expected outcome instead of forcing a margin value.

Worked record examples

These examples use the same row types as the calculator body. The useful comparison is the calculator status plus the action that should remain with the job record.

Worked record examples
Work settingEntered rowCalculator statusValues to keep with the record
Switchboard handover`RCD-1`, 30 mA, rated-current row, Type 2 300 ms basis, 28 ms measured.Within entered criterion.30 mA test current, 300 ms maximum, 272 ms margin, board schedule and instrument record.
Temporary-board register`RCD-2`, 30 mA, non-operating row, 5 s duration, no trip.Within entered criterion.15 mA test current, 5 s no-trip duration and the selected test point.
Maintenance review`RCD-5`, push-button function did not operate.Function test failed.Device reference, function-row wording and open device/procedure review action.
Contractor check`RCD-3`, 30 mA, five-times row, manual 40 ms maximum, 18 ms measured.Within entered criterion.150 mA test current, 22 ms margin and the source of the manual 40 ms basis.

If a row is flagged, do not change the row type or criterion until the status looks better. First confirm whether the original record, the test setup or the device itself needs correction.

Retest or investigate

Do not solve a flagged row by changing the inputs until the status looks better. Keep the calculator record aligned with the measurement in front of the user. If the record is wrong, correct the record or retest. If the result repeats, investigate the device, test point, circuit state, connected loads, instrument setup and applicable procedure.

Status action matrix
Calculator statusRecord wording to useFollow-up focus
Within entered criterionKeep the row with its device reference, row type, criterion basis and original instrument record.Confirm the basis is the intended one before carrying the row into handover or maintenance records.
Unexpected trip recordedState that the non-operating-current row tripped.Confirm test setup and investigate if the trip repeats under controlled conditions.
No trip recordedState that a timed row recorded no trip.Confirm supply condition, test connection, device condition and the applicable procedure.
Above entered timeState the measured time and entered maximum.Check transcription, criterion basis, test point and manufacturer data before retesting or escalating.
Function test failedState that the push-button function did not operate.Keep it as a function-check finding and review the device or procedure.

This distinction keeps the worksheet from becoming a substitute for professional judgement. It also helps the exported row say something useful. Repeated no trip on timed row, device and setup to be investigated is clearer than a bare status.

RCBO record boundary

For RCBOs, keep the same residual-current test-row language but make the device reference specific enough to identify the RCBO, circuit and board. The overcurrent side of the RCBO, prospective fault current, loop impedance and cable withstand are separate protection questions handled by other checks.

Using the export in a job pack

The PDF export should be treated as a working record, not the original evidence by itself. Keep the original instrument record, the board or circuit schedule, the device label and any site procedure with it. The calculator output is useful when it gives the supervisor or future maintainer a compact summary of what was entered and why the row was flagged or kept for the job file.

For a clean timed row, identify the board, device, test point and criterion basis. For a flagged row, write the action still open. Do not use the export as final verification; it compares one recorded row with the basis entered by the user.

If the RCD or RCBO row raises a wider protection concern, move to the specific calculation rather than overloading this page. Use fault-loop impedance for loop-current context, short-circuit current for prospective fault-current context, and I2t cable withstand for conductor withstand against entered fault energy.

Source basis and Australian boundary

Australian RCD and RCBO testing records sit inside a wider verification and workplace-safety context. Public Australian guidance commonly distinguishes push-button checks from other testing records, and public guidance for some contexts describes no-trip and trip-time rows such as half-current non-operating checks, rated-current checks and Type 1 or Type 2 timing examples. This calculator uses those public concepts only as worksheet context.

The controlled requirements still belong to the current standards, regulator guidance, workplace procedure, manufacturer instructions and the competent person responsible for the work. AS/NZS 3760, AS/NZS 3000 and AS/NZS 3017 are mentioned as context because they are part of the Australian testing and verification environment. This page does not reproduce controlled tables and does not claim that the preset values are universal for every device or site.

Local authority, DNSP and site requirements can also matter. They may affect the test environment, documentation required before energisation or handover, acceptable evidence for a workplace record, and escalation path for a failed result. Treat those requirements as outside the calculator and attach the relevant basis to the job record where it affects the row.

Manufacturer instructions can override a generic worksheet assumption as well. Device type, electronic RCD behaviour, test button operation, line/load connection requirements, nuisance-trip history and replacement instructions should be checked from the product documentation or a competent person before relying on the row.

Stop points

  • The device reference cannot be matched to the board schedule, circuit list, asset register or test sheet.
  • The row type is uncertain or the criterion basis is not known.
  • The measured value is missing from a timed row that should have a measured time.
  • The test point, supply condition, connected load state or instrument setup cannot be confirmed.
  • A non-operating row trips, a timed row records no trip, a timed row exceeds the entered basis, or a push-button check fails.
  • Manufacturer instructions, workplace procedure or local requirements specify a different basis from the value entered here.
  • The calculator output is being used as a compliance certificate, replacement test record or device replacement decision.

Temporary board register

A construction-board register records a 30 mA device held closed for 5 s at the non-operating-current row.

Device reference
DB-RCD-01
Rated residual current
30 mA
Test row
Non-operating current
Criterion basis
5 s no-trip duration
Observed outcome
No trip
No-trip duration
5 s
  1. Test current30 mA x 0.5 = 15 mA
  2. No-trip rowRecord expects no trip for 5 s at the non-operating-current row.
RCD recordWithin entered criterion

The entered row is readable as a worksheet comparison.

This is a register-quality row only if the device reference, test point and instrument record identify the same RCD.

  • The register identifies the tested device and location.
  • The no-trip row was intentionally selected for this record.
  • The project procedure still governs record acceptance.

Switchboard handover record

A switchboard handover pack includes a Type 2 timed row with a 28 ms measured trip time.

Device reference
DB-RCD-04
Rated residual current
30 mA
Test row
Rated-current trip time
Criterion basis
Type 2 300 ms
Observed outcome
Trip recorded
Measured trip time
28 ms
  1. Test current30 mA x 1 = 30 mA
  2. Time comparison300 ms maximum - 28 ms measured = 272 ms margin
RCD recordWithin entered criterion

The entered row is readable as a worksheet comparison.

The row is suitable as a worksheet comparison when the entered 300 ms basis is the intended project record basis.

  • The board schedule and device reference match.
  • The 300 ms basis is entered deliberately for this row.
  • The original instrument record remains with the handover file.

Maintenance review flag

A facilities maintenance record shows the push-button control did not operate the device.

Device reference
DB-RCD-07
Rated residual current
30 mA
Test row
Push-button function
Criterion basis
Function outcome
Observed outcome
Did not operate
  1. Function checkRecorded outcome is did not operate, so the row remains a function-check item.
RCD recordFunction test failed

The record should be held for device, test setup or procedure review.

The row should be treated as an action flag for device or procedure review, not as a timed operating-current result.

  • The push-button row is recorded as a function check.
  • The maintenance record separates this result from any timed test.
  • Device review happens outside the calculator.

Five-times trip row

A switchboard record lists a 30 mA RCD on a five-times row with an entered 40 ms maximum and an 18 ms measured time.

Device reference
RCD-3
Rated residual current
30 mA
Test row
Five-times trip time
Criterion basis
Manual 40 ms
Observed outcome
Trip recorded
Measured trip time
18 ms
  1. Test current30 mA x 5 = 150 mA
  2. Time comparison40 ms maximum - 18 ms measured = 22 ms margin
RCD recordWithin entered criterion

The entered row is readable as a worksheet comparison.

The row is useful only when the entered 40 ms basis matches the project, workplace or manufacturer record for that five-times test.

  • The five-times row is deliberately selected.
  • The 40 ms maximum is the entered comparison basis for this record.
  • The original instrument result remains available for review.

Questions

Does this calculator certify the installation?

No. It compares one entered row with the entered criterion and record basis. Final reliance still depends on the test instrument, device condition, workplace procedure and current Australian requirements.

Is a push-button test enough on its own?

No. The push-button row is a function check only. It does not replace a timed row or a competent person's record review.

Why do Type 1 and Type 2 labels appear?

They are entered criterion labels for the timed row. They help the user record a 40 ms or 300 ms basis when that is the comparison being used.

Can the criterion be entered manually?

Yes. The manual criterion field allows a project, workplace or manufacturer time basis to be recorded when the entered row needs a different threshold.

What if the device trips on the non-operating-current row?

That is an unexpected trip and needs follow-up outside the calculator. The row should not be read as a satisfactory no-trip record.